Description of the Experiment

Figure 1: Photo of the experimental setup

Plasmas have different characteristics from other states of matter and many diagnostic tools have been developed in order to measure their parameters . This experiment allows the measurement of some of these characteristics using one of the most simple methods, an electrical probe, the Langmuir probe. It consists of a thin filament made of conductive material placed inside the plasma, which either attracts or repels the electrons in the plasma in accordance with its polarization. Measuring the probe I-V characteristic, that is the relationship between the polarization voltage and the respective collected current, you can determine the electron temperature and density of the plasma.

Links

  • Video: rtsp://elabmc.ist.utl.pt/langmuir.sdp
  • Laboratory: Advanced in e-lab.ist.utl.pt[1]
  • Control room: Langmuir Probe
  • Level: ****

Videos

Experimental Apparatus

The vacuum chamber consists of a Pyrex cylinder on which a glow discharge is produced by RF power. A tungsten filament with a diameter of 0.2mm is placed in the middle of the chamber partially isolated by an alumina cover and only a 10mm tip is exposed to the plasma, acting as a Langmuir probe. This central filament has a ground reference around the probe with approximately 200mm, in order to be much longer than the probe itself.

The RF generator (3.5W, 50kHz, 1kV) that ionizes the gas and generates the plasma is connected to two parallel molybdenum plates similar to a capacitor.

NOTE: Gas pressure correction should be computed according Pfeifer's vacuum gauge PPT100 spreadsheet.

Protocol

Determination of the electron temperature and density

Figure 2: Raw Data of the experiment
Figure 3: Linear fit to the ion saturation region
Figure 4: Fit to probe characteristic equation

The Langmuir probe uses one or more electrodes placed inside the plasma to measure the current collected while applying a voltage. This will provide the I-V characteristic for that plasma. Normally a triangular or sawtooth wave is used to get equi-separated points.

When the probe is electrically isolated, a plasma sheath is formed in the interface between the plasma and the probe. To compensate for the higher mobility of the electrons, the probe will attain a floating potential,\(V_f\), negative in relation to the plasma potential itself, \(V_p\). The density at the sheath entrance is roughly half of the density far away from the probe.

The probe voltage, \(V_s\) is sweeped in relation to the ground winding filament, using a variable voltage source. If the polarization of the probe compared to the plasma potential is negative enough, all the electrons will be repelled and the ion flux to the probe is independent of the potential applied. In a totally ionized plasma, this ion saturation current is described by the following expression: [math] i^+_{sat} = j^+_{sat} A_s = \frac{e \, n \, c_s \, A_s}{2} [/math]

Being \(j^+_{sat}\) the ion current density, \(A_s\) the contact surface of the probe, \(e\) the electron charge, \(n\) the ion density in the plasma, and \(c_s\) the ion sound speed.

When the polarization applied to the probe is increased, the voltage drop in the sheath is reduced and the electrons will be able to reach the probe. Taking a Maxwell distribution of the speed of the electrons, the relation between current and tension will be: [math] i = -i^+_{sat} \left( 1- e^{\dfrac{e}{k T_e} (V_s - V_f)} \right) [/math]

Being \(T_e\) the electron temperature. This expression assumes that there is only one probe and that it is non-perturbative.

From the last equation it is possible to extract an estimate of the floating potential, \(V_f\). This is done by taking the value at which the characteristic crosses zero.

It will also be easily seen that the data does not follow the regular characteristic on the ion saturation side, it should be constant instead of having a slope. This has to do with the fact that the sheath thickness expands with the applied voltage. To correct this, a linear function is fitted on this side and the slope is subtracted (as seen in Figure 3). Alternatively you can include this contribution in the equation:

[math] i = -i^+_{sat} \left( 1- \alpha\ (V_s-V_f) - e^{\dfrac{e}{k T_e} (V_s - V_f)} \right) [/math]


However, you want to get the density, so you have to make another correction, which corresponds to adding the value of current at the point where you know the current is completely due to ions, namely to voltage values much lower than the floating potential (in this case you use two times the floating potential, hence the need for our initial estimate).

After that, the experimental data has to be adjusted to the equation presented above (as seen in Figure 4). From the fit the \(T_e\) can be extracted, as well as the ion saturation current, \(i^+_{sat}\) and the floating potential \(V_f\), which should have a value close to the one we determined empirically.

By knowing the area of the probe and the speed of sound in the ions of the plasma, you can determine the electron density in the plasma. Since \(c_s = \sqrt{\dfrac{k T_e}{M}}\), and given that the probe is \(10mm\) and has a diameter of \(0.2mm\) you can use the \(i^+_{sat}\) to determine the density.